PRODUCTS – FIBRE OPTICS – SERIES 2000 POWER METER
The Series 2000 Fibre Optic Power Meter is a desktop unit. Versatile and modular,
it is standard equipment with the Benchmark OFS System or can be purchased as an optional extra for greater flexibility
and expandability of existing configurations.
Features
- Versatile microprocessor based instrument with keypad selection and LCD
status display
- High performance optical power meter for precision measurement on multi-mode
and single-mode fibre optic links
- Modular design with user-pluggable detector modules to cover different
ranges of input wavelengths
- Connector adapter provision for mating with widely used fibre-optic connectors
such as SMA, FC and ST
- Wide wavelength range covered by each detector module
- Completely calibrated over entire input wavelength range
- Wide dynamic range : +3 to -60 dBm
- Absolute and relative Power measurements
- Readout in both dBm and mW/µW/nW (Auto ranging)
- Store and recall up to 10 values
- Hold facility for snapshot readings
- Highly accurate and stable operation (0.15dB) with high resolution (0.02dB)
- Analog electrical output available
- Choice of averaging speeds
- RS-232C interface (optional) for external control and data logging through
PCs
Specifications
Functional
 |
|
|
|
|
 |
Max number of sources |
 |
Two |
 |
| |
Source selection |
|
All OFF, one ON, all ON |
|
| |
Keypad |
|
6 x 2 membrane keypad |
|
| |
Key Functions |
|
Enter, Reset, Up & down Arrows, Hold, Store, Recall,
Relative measurement Selection of: dBm/Watt display, Averaging speed, Calibration
wavelength and Local/Remote control |
|
| |
Display |
|
1) 20 X 2 dot matrix LCD for status display ii) 4 digit
LCD for power measurement |
|
| |
Status indication |
|
Power unit (dBm/microW), relative measurement, hold, indication,
calibration wavelength, stored value index,averaging speed, local/remote
control |
|
| |
Display mode |
|
dBm/mW/microW/nW (auto-ranging in Watt scale) |
|
| |
Display resolution |
|
0.02 dBm |
|
 |
|
|
|
|
Output
 |
|
|
|
|
 |
Analog signal output |
 |
Log output (BNC socket, 1K Ohms impedance |
 |
 |
|
|
|
|
Interface
 |
|
|
|
|
 |
Communication interface (optional) |
 |
RS-232 [EIA data lead only] |
 |
 |
|
|
|
|
Environmental
 |
|
|
|
|
 |
Operating temperature |
 |
0o to 40o C |
 |
| |
Storage temperature |
|
-10o to 70o C |
|
 |
|
|
|
|
Physical
 |
|
|
|
|
 |
Dimensions - width x depth x height |
 |
320mm x 360mm x 135mm |
 |
| |
Weight |
|
4kgs |
|
| |
Power consumption |
|
20W max |
|
 |
|
|
|
|
Photo Detector Modules
*Choose any one while ordering
Ordering Information
Series 2000 Optical Power Meter
 |
Description |
|
Part # |
|
 |
Series 2000 Optical Power Meter Mainframe & PM2011 -XX
Silicon Photo Detector Module |
 |
PM 2001 -XX |
 |
| |
Series 2000 Optical Power Meter Mainframe & PM2012 -XX
Germanium Photo Detector Module |
|
PM 2002 -XX |
|
| |
Series 2000 Optical Power Meter Mainframe , PM2011 -XX Silicon
Photo Detector Module & PM2012 -YY Germanium Photo Detector Module
|
|
PM 2003 -XX-YY |
|
| |
PM2011 Silicon Photo Detector Module |
|
PM 2001 -XX |
|
| |
PM2012 Germanium Photo Detector Module |
|
PM 2002 -XX |
|
| |
XX or YY = SM for SMA adaptor cap; FC for FC adaptor cap;
ST for ST adaptor cap
|
|
|
|
 |
|
|
|
|
*To be specified while ordering